|
|

CMA Homepage
Information for users
Reservations
Mass Spectrometry
NMR
Photon Spectroscopy
Current Rates
Sample Drop-off
Useful Links
Mark E. Bier
Director, CMA
mbier@andrew.cmu.edu
412-268-3540 (office)
412-268-6897 (fax)
Roberto R. Gil
Director, NMR Facilities
rgil@andrew.cmu.edu
412-268-4313(office)
Carnegie Mellon University
Department of Chemistry
Mellon Institute Bldg.
4400 Fifth Avenue
Pittsburgh, PA
15213-2683
|
|
Mass Spectrometry
|
|
|
|
The CMA has
three mass
spectrometers available for general use:
Applied Biosystems Voyager DE-STR MALDI-TOF
Thermo LCQ ESI/APCI
Ion Trap
Waters Q-TOF II ESI/APCI/Nanospray
|
|
|
General Usage Guidelines:
-
For users
expecting to have fewer than 3 samples per month, please use our
sample drop-off service.
-
Users
expecting to run samples frequently (more than 3/month for a year or
longer) should
contact Dr. Mark
Bier to schedule training on the required instrument(s).
-
CMA provides
a sample preparation bench for general use with common resources such as
pipettes, sample vials etc. but users must provide:
-
Solvents
for dilution, instrument rinsing etc.
-
MALDI:
Calibrant, matrix, sample plate (Part # V700666 from
Applied Biosystems).
-
LCQ/Q-TOF:
Square-tipped ("LC-style") gas-tight syringe.
-
Users must
reserve time on the
instrument
calendar prior to use.
-
For more
information about mass spectrometry, see our useful
links.
|
|
|
|
Applied Biosystems Voyager DE-STR MALDI-TOF
Funded
by NSF grant #CHE-9808188
|
|

The PerSeptive Voyager STR
MS is a matrix assisted laser desorption ionization (MALDI) time of flight
(TOF) mass spectrometer. Ions are formed in the gas phase from a laser
focused
onto a matrix/analyte solid sample. The matrix compound is typically a
small organic molucule, chosen for its ability to absorb photons at 337 nm
from the nitrogen laser. Gas phase ions (positive ions are typically
protonated) are then accelerated down the 2 meter flight tube with 20-25
keV of kinetic energy (KE). Low m/z ions hit the detector first followed by
the slower, higher m/z ions. To a first order, the ions follow the equation
KE=1/2mv2. Typical analyte molecules include polymers, peptides,
proteins, and DNA.
See how MALDI
works!
System Overview:
-
Mass range 1-400,000 Da
-
Positive and negative ion
modes
-
Linear (R=3000) and reflector
(R=15000) modes
-
Post-source decay
(PSD) with CID
-
337nm nitrogen
laser
-
100 well standard sample
plate (user-supplied)
-
Polymeric analysis tool (PDI, Mn, Mw, Mz
calculation)
-
5ppm
mass accuracy in reflectron mode with internal reference
-
Femtomole-level detection with proper sample preparation
Calibration:
Sample Preparation:
-
Sample and matrix should be
prepared in similar or miscible solvent systems.
-
Preparation typically only
requires 1無 of sample
-
Selecting the proper matrix
has a large effect on results. This decision can be based on previous
successful MALDI analyses of similar samples, the
NIST polymer
recipes page for polymer samples, or see the general guide found in the Voyager manual (section
3.1.1 - "Selecting a Matrix").
-
Samples must be completely dry
before loading into the instrument.
-
For positive ion samples, a
small amount of acetic acid or TFA (0.1% v/v) can be added to enhance the signal.
-
Approximate recommended sample
concentrations are as follow:
Back to Top
|
|
|
|
Thermo-Fisher
LCQ ESI/APCI Ion Trap
Funded by NSF grant #DBI-9729351
|
|

The Finnigan LCQ is a
quadrupole field ion trap mass spectrometer with electrospray ionization
(ESI) and atmospheric pressure chemical ionization (APCI) sources. The LCQ
has MSn capabilities, which is a powerful analytical tool for
use in structural analysis. This mass spectrometer can provide molecular
weight and sequence information for small peptides (MW below 2000 Da) at
the femtomole level with proper analyte introduction.
System Overview:
-
Optimum mass range 50-2000 m/z, 4000 m/z
in limited capacity
-
Resolution of >1000 across optimum mass range on a full ms scan
-
Zoom
and turbo scan features
-
Positive / negative ion modes
-
ESI and APCI sources
-
MSn (i.e. MS-MS-MS-...) mode for
structural analysis
-
LC-MS
capability
-
Protein
identification with SEQUEST software
-
Deconvolution tool for protein mass measurement
Calibration:
Sample Preparation:
-
All samples
should be in an ESI-friendly solvent system such as 50/50 Water/Acetonitrile,
Water/Methanol, etc.
-
For positive
ion samples, a small amount of acetic acid (0.1% v/v) can be added to
enhance the signal. TFA is not recommended.
-
For direct
injection, prepare at least 100無 of sample. For flow-injection
analysis 20無 is typically sufficient.
-
Recommended
sample concentrations are as follow:
Back to Top |
|
|
|
Waters Q-TOF II ESI/APCI Quadrupole-TOF |
|

The Micromass Q-TOF II is a hybrid mass spectrometer designed specifically
for high performance in the mass range 100-2000 m/z, especially with protein
identification in mind. Samples are introduced via ESI, APCI, or
Nanospray ionization sources, analyzed (optionally) by the quadrupole,
fragmented (optionally) in the hexapole collision cell, then accelerated
orthogonally into the TOF tube for high-resolution mass analysis. The
Q-TOF II is capable of MS and MS-MS analysis and achieves resolution
exceeding 10,000. It can also be used to detect ions above 2000 m/z
although not specifically optimized to do so.
System Overview:
-
Optimum mass
range 2-2000 m/z (theoretically 2-32,000 m/z)
-
Quadrupole and
orthogonal V-reflectron TOF analyzers
-
Positive /
negative ion modes
-
ESI, APCI, and
nanospray ion sources
-
Auto-injector
valve for flow-injection analysis
-
Z-spray source
inlet configuration for reduced noise & contamination
-
MS and MS-MS
(CID) modes
-
Resolution
exceeding 10,000 within optimum mass range
-
Femtomole-level
sensitivity within optimum mass range
-
5ppm mass
accuracy with proper internal reference over 150-900 m/z
Calibration:
Sample
Preparation:
-
All samples
should be in an ESI-friendly solvent system such as 50/50 water/acetonitrile,
water/methanol, etc.
-
For positive
ion samples, a small amount of acetic acid (0.1% v/v) can be added to
enhance the signal. TFA is not recommended.
-
For direct
injection, prepare at least 100無 of sample. For flow-injection
analysis 20無 is typically sufficient.
-
Approximate
recommended sample concentrations are as follow:
Back to Top |
|
|